Raytek A-T40-WC Air/Water-Cooled Housing

Rp14.915.000

Out of stock

Category:

Description

The air/water-cooled housing allows the sensor to be used in ambient temperatures up to 120°C (250°F) with air
cooling, and 180°C (356°F) with water cooling. The cooling media should be connected using 1/8” NPT fittings
requiring 6 mm (0.24 in) inner diameter and 8 mm (0.31 in) outer diameter for the tube. Airflow should be 1.4 to
2.5 l/s (0.37 to 0.66 gallons/s) at an air temperature of 25°C (77°F). Water flow should be approximately 1.0 to
2.0 l/min (0.26 to 0.52 gallons/min) at a water temperature between 10 and 27°C (50 to 80.6°F). Chilled water
below 10°C (50°F) is not recommended.

The Air/Water-Cooled Housing is made from stainless steel and is delivered with air purge. The air purge should be always used to keep the lens dry.

Note
For ambient temperatures exceeding 175°C (350°F), the ThermoJacket can be used. This accessory allows
operation at ambient temperatures up to 315°C (600°F)!